Complete UV/VIS/NIR Systems
Liquids and Gases
Worldwide Sales and Support
Liquids and Gases
Worldwide Sales and Support
![]() |
The Guided Wave Model 412 process Analyzer offers remote multi-channel near-infrared (NIR) or visible (Vis) spectroscopic scanning and analysis. Essentially a smart chemical sensor, the Model 412 provides exceptional signal-to-noise, excellent wavelength stability, NIST traceable wavelength calibration, dual-beam optics and built-in diagnostics. Process operators, engineers, and researchers will appreciate the Model 412's ability to monitor up to 12 process streams (or points within a stream) with accuracy, repeatability, and reliability.
The Model 412 is the core of a comprehensive process analyzer system that includes the spectrometer; one or more NIR probes, fiber optic cables, and intelligent scanning-and-analysis software. The Model 412 uses near-infrared radiation to collect spectral data from liquids, gases, slurries, glass, and polymer-based films. The spectral data are interpreted by the analyzer system's software, to determine the composition and/or physical characteristics of the material.
In remote spectroscopy the analyzer transmits radiation to one or more probes installed directly in your process reactors or pipes.
The Model 412's dual-beam design, coupled with its high output scanning grating, puts more light in the fiber; providing the highest available signal-to-noise ratio. Every Model 412 also provides excellent wavelength accuracy (NIST traceable) and superb wavelength repeatability.
Here are just a few of the many applications for which our customers rely on the Model 412 every day:
|
Model |
412 VIS |
412 NIR standard
|
412 NIR extended
|
|
(rated) |
380 - 1070 |
800 - 1700 |
1000 - 2100 |
|
|
400 - 1000 |
900 - 1600 |
1050 - 2000 |
|
Wavelength Accuracy, nm |
±0.28 |
±0.20 |
±0.20 |
|
Wavelength Repeat., nm |
±0.02 |
±0.02 |
±0.02 |
|
Wavelength Stability (rms/24hrs) |
±0.05 |
±0.05 |
±0.05 |
|
Minimum Step Size, nm |
0.2 |
0.2 |
0.2 |
|
Spectral Resolution, nm |
3.2 |
6.8 |
6.8 |
|
Photometric rms Noise |
< 20 µAU |
< 13 µAU |
< 26 µAU |
|
Photometric (Baseline) Stability (rms/24hrs) |
< 1 mAU |
< 1 mAU |
< 1 mAU |
|
|
4.0 decades |
4.0 decades |
4.0 decades |
|
Optical Connections |
SMA 905; 500 µm Diameter Single Strand Fiber Optic Cable |
||
|
Data Communications |
RS422, RS-232 (DB9 Null Modem Cable), or optical fiber modem |
||
|
Dimensions |
30” (w) x 17.3” (d) x 41.3” (h) [76 cm x 44 cm x 105 cm] |
||
|
Weight |
200 lbs [91 kg] |
||
|
Power |
1500 W, 110/220 Vac, 50/60 Hz |
||
|
Climate (Temperature/Humidity) |
0C to 40°C Shaded 0-90% non-condensing
|
||
The Guided Wave Model 412 comes installed with an air-cooled internal climate control system for General Purpose Environments. This instrument can also be packaged for hazardous environments. These options include Class I Division 2 groups C & D and CENELEC EEx pdem [ib] II C T4.
The Model 412 must be used with an ICU (Instrument Control Unit, a PC-compatible computer) that meets or exceeds the following specification:500 MHz or better CPU; 512M RAM, DVD/CD-ROM, open slot for communication card, Windows XP or Vista OS.