Complete UV/VIS/NIR Systems
Liquids and Gases
Worldwide Sales and Support
Liquids and Gases
Worldwide Sales and Support|
Probe 2 |

|
Enclosures |
|
|
Dimensions |
|
|
Certifications |
|
|
Purge |
Expo-Safety Systems, continuous flow sub-miniPruge (1ZCFSS1S). Intrinsically safe switch closure and pneumatic "winkie" fo purge loss. 1/2" NPTF purge fitting. Purge vent orofice is Size 1 (for 0.4 cfm; 10 L/min). Allow 6 min. purge cycle (min. 4 volumes) prior to power up |
|
Purge Gas |
110 p.s.i. clean, dry compressed air or inert gas |
|
Fiber connections |
SMA 905 400 - 600 micron (core) |
|
Detectors |
Max. 2+2, 3+3, 4+2 for Group "A" + "B"
|
|
Filters |
10 nm FWHM for UV/VIS, 15 nm for NIR typical. 4 OD out-of-band blocking |
|
Light Source |
|
|
Wavelength Drift |
0.1 nm with Si detector 0.2 nm near 1450 nm with InGaAs detector |
|
Photometric Drift |
<2 mAU with ±5ºC variation (±2 to 3 mAU with Xe) 0.2 mAU/ºC at 0 AU typical <2 mAU over 2 weeks |
|
Photometric Noise |
<1 mAU (3 mAU with Xe source) |
|
Outputs |
4 4-20 mA isolated, self-powered. 2 4-20 mAU additionally optional for a total of 6. |
|
Environmental |
10 - 40 ºC, 10-90% rel. humidity (non-condensing) |
|
Power |
110 - 240 V AC, 50-60 Hz, 70 VA (24 V DC, 2.1 A |