ClearView SMART PHOTOMETER



For North America
Class 1, Div. 2 Groups B-D 
For Europe
ATEX  II  2 G EEx d IIC T6

Optical Solutions' brand ClearView photometer was originally designed to provide a low-cost means to measure color and turbidity using the flexibility of optical fibers to separate the analyzer from the sample cell or probe.  This had two advantages:
  1. The detector and electronics are not in contact with potentially hot pipes, as in site glass turbidity or color meters
  2. The probe can be removed and calibrated with a small amount of sample where site glass systems have no easy means of validating
Turbidity and Color
With a special fiber optic turbidity probe, ClearView is able to quantitatively measure turbidity in the 0-900 NTU range using 90 degree off-axis light scattering inside large pipes and reactor vessels.  The light beam is shown here crossing the sample space. The small optic below the beam picks up the scattered light and ClearView converts the signal into NTU.


The initial emphasis on APHA color was extended to the measurement of fuel color in refineries and pipelines, including:

  • Saybolt color in aviation fuel
  • ASTM color in diesel fuel or gas oil
  • Red dye concentration (ppm) for non-automotive diesel to comply with U.S. Internal Revenue Service taxation requirements

The picture above is of an insertion probe contained in a safe retracton device measuring fuel color in France.

ClearView takes advantage of the patented ChemView technology with multiple detectors, each having a narrow (10-15 nm) optical filter selected for the wavelength of interest.  As a result, we can place a Silicon detector tuned to APHA color at 450 nm in the visible (VIS) adjacent to an InGaAs detector tuned to water conentration at 1440 nm in the near-infrared (NIR) and measure them simultaneously.  Therefore, ClearView evolved from just measuring color and turbidity at VIS wavelengths, to measuring composition in the NIR.
 
  • Simultaneous measurement of color and % water
  • Simpler NIR measurements
    • styrene in benzene
    • % water in organics
    • %water in acids
    • solvent mixtures (2-3 components)
    • ethylene or propylene gas
 
ClearView will NOT measure
  • UV  (<390 nm) 
  • NIR wavelengths > 1650 nm (e.g. 100 ppm water)

In addition, ClearView compensates for:
  • Sample temperature (needed for NIR water measurements)
  • Sample pressure (needed for UV or NIR gas measurements)
by receiving RTD temperature or pressure transducer signals. Thus, you do not have to correct the readings in your process control computer.

The operator will mostly see the following screens on ClearView, indicating how easy the system is to use.  Enter the coefficients from the calibration, display the predicted conenctrations for color, scale the 4-20 mA analog outputs and continuously monitor your process.

MEASUREMENT MODE
NTU               2.9
SAYBOLT    12.7
 
 
 
 
 
CONF
DIAG
ADJ
ABS
Measurement Mode Screen displays all answers
 
 
 
 
 
ABSORBANCE VALUES
5975                  2177 MV
455                     98 MAU
 
         REFERENCES
670                 3134 MV
 
 
ZERO
CLR
HELP
EXIT
Absorbance Screen shows the absorbances at all wavelengths used in the Coefficient Screen to determine chemical concentration. The reference voltage is a key diagnostic.
 
MAIN CALIBRATION
 
NAME: APHA GROUP A
 
A0 = +0.00000E+00
T1 = +1.23450E-01
A2 = +0.34500E+00 455NM
RF = +0.00000E+00 700NM
 
 
 
 
 
MAIN CALIBRATION
 
NAME: SAYBOLT GROUP A
 
S0 = +0.00000E+00
S1 = +1.23450E-01
S2 = +0.34500E+00
S3 = +5.00000E-01
S4 = +0.00000E+00
 
 
 
 
 
Coefficient Screen allows you to input coefficients for each wavelength, including sample temperature, to convert absorbances to chemical concentrations. Two types of calibrations are shown here. The upper screen shows a linear calibration with sample temperature compensation. The lower screen shows a polynomial calibration typically used for Saybolt color.
 
 
 
NTU
MA
 4 = +0.0000E+00
20 = +2.0000E+01
 
 
 
INC
DEC
HELP
EXIT
Analog Output Screen scales your 4-20 mA output in concentration units for each answer.
 
 

 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 


 



Specifications

Number of detectors

Up to 4; 1 reference, up to 3 analyzing wavelengths

Lamp

Patented Optical Solutions' StabLamp with optical feedback,  precision-mounted Tungsten-Halogen bulb; ~8 months bulb life nominal

Photometric Drift

<2 mAU over 2-3 weeks with ±5 ºC variation

Wavelength Drift

0.1 nm VIS typical

0.2 at 1450 nm

Dynamic Range

3.6 AU typical near 1450 nm (non-linearity <1% at 1.6 AU)

Enclosures

Adalet meeting C1D1 Groups B-D

CorTem meeting ATEX II 2 G EEx d IIC T6

Outputs

2 self-powered, isolated 4-20 mA standard, 3 optional, 500 ohm load max. Two user-programmable opto-closures. One opto-closure pre-programmed for any fault.  Large  3 x 4" LCD display controlled by 3 external buttons (C1D1) or magnetically activated pen (ATEX)

Inputs

One programmable discrete;  One input for RTD or pressure transducer to correct for sample temperature or pressure variations.

Fiber Optics

SMA 905 connectors standard

400-600 micron core fibers

Environmental

0 - 50 ºC; 10-90% RH,  non-condensing

Dimensions

C1D1 8" x 8" x 8.5" 20lbs 
(20 cm x20 cm x22 cm, 9 kg) 

Power

24 VDC, <1A, (AC-DC converter supplied with unit for lab testing)




 


 

.
© 2006 Guided Wave, Inc.