HeaterCell for Temperature Control of Vials and Cuvettes


  • Melt samples that are solid at room temperature
  • Establish calibrations at melt temperatures
  • Include the effect of temperature in calibration models of liquid mixtures
  • Use precision cuvettes or disposable vials with dual-optics selection
  • Room tempeature to 280 ºC
  • SMA fiber optic connections


HeaterCell is a fiber optic accessory designed to heat glass cells up to 280 ºC.  You can melt polymers to temperatures found in the reactor.  You can also keep samples at a constant temperature for calibration.  Many calibrations involving water and alcohols are temperature dependent.  By analyzing a sample set over the range of expected temperatures, HeaterCell can provide calibrations that take the effect of sample temperature into account in Guided Wave spectrophotometers and photometers. 

HeaterCell was designed with dual optics that are accessed with the lever on the right side of the device.  Collimated light is used for square or rectangular cuvettes.  Convergent light is used for disposable vials.

HeaterCell has SMA connectors on each side that connect to our spectrophotometer or photometer.

Specifications



Dimensions 140 H x 226 W x 140 D (mm)
5.5 H x 8.9 W x 5.5 D (inches)
Weight 5 kg (11 lbs)
Power (110V operation) 90-135VAC at 0.4A max; 50-400 Hz
(220V operation) 180-250VAC at 0.2A max; 50-400 Hz
Fusing:  (3AG)  0.5A  250VAC
Operating
Environment
0 to 50*C (32 to 122*F)
10 to 90% Relative Humidity (0 to 40*C)
10 to 50% R.H. (at 50*C) non-condensing
Cell Temperature  Range Ambient to 280 ºC
Temperature Accuracy ±1.0*C from ambient to 100 ºC
Fiber Diameter < 1 mm with a Numerical Aperture of 0.2
Optical Throughput >50% transmittance referenced to coupled fibers (350-2000 nm)
Thermal Protection Self-limiting at high temperature


.